After data collection resumed following an extended outage on 3/20/07, several 
housekeeping data fields failed maximum value checks. It was determined that a power event occurred 
on this day, disabling housekeeping board 2 of 3 in the SCE box. 

Scientifically relevant data continue to be valid.

Disregard the housekeeping values when looking at data after 3/20/07:
Suggestions: 
Scientifically relevant data are valid
Measurements:sgpaeri01ch1C1.a1:
  • airNearBBsTemp
  • coolerExpanderTemp
  • BBcontroller2temp
  • fixed97KohmResistor
  • interferometerSecondPortTemp
  • BBcontroller1temp
  • mirrorMotorTemp
more
sgpaeri01summaryC1.a1:
  • BBcontroller1temp
  • interferometerSecondPortTemp
  • BBcontroller2temp
  • mirrorMotorTemp
  • coolerExpanderTemp
  • airNearBBsTemp
more
sgpaeri01ch2C1.a1:
  • interferometerSecondPortTemp
  • coolerExpanderTemp
  • mirrorMotorTemp
  • BBcontroller2temp
  • BBcontroller1temp
  • airNearBBsTemp
  • fixed97KohmResistor
more
sgpaeri01engineerC1.a1:
  • fixed97KohmResistor
more