Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/27/2011 | 0000 | 05/03/2014 | 2359 | Does not affect quality |
03/12/2011 | 0000 | 03/16/2014 | 2359 | Does not affect quality |
Subject: | TWP/KAZR/C1/C3 - Reprocessing Notice (RID 1118) |
DataStreams: | twpkazrhiC3.a1, twpkazrgeC3.a1, twpkazrauxC1.a0, twpkazrspeccmaskgecopolC3.a0, twpkazrspeccmaskmdcopolC3.a0, twpkazrgeC1.a1, twpkazrspeccmaskhixpolC3.a0, twpkazrspeccmaskhicopolC1.a0, twpkazrspeccmaskgecopolC1.a0, twpkazrspeccmaskhicopolC3.a0, twpkazrspeccmaskmdxpolC3.a0, twpkazrspeccmaskgexpolC3.a0, twpkazrauxC3.a0, twpkazrhiC1.a1 |
Description: | See below for details of reprocessing that was completed on 20150626. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- The wrong calibrations were used due to a problem with the software. Mentor discovered the reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed. |
Suggestions: | |
Measurements: | twpkazrspeccmaskgexpolC3.a0: more
twpkazrspeccmaskhicopolC3.a0: more twpkazrgeC3.a1: more twpkazrspeccmaskhicopolC1.a0: more twpkazrauxC3.a0: more twpkazrspeccmaskhixpolC3.a0: more twpkazrauxC1.a0: more twpkazrspeccmaskmdcopolC3.a0: more twpkazrgeC1.a1: more twpkazrhiC3.a1: more twpkazrspeccmaskgecopolC1.a0: more twpkazrspeccmaskmdxpolC3.a0: more twpkazrspeccmaskgecopolC3.a0: more twpkazrhiC1.a1: more |