DQRID : D140307.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/21/2013045103/03/20141659Suspect
more
Subject:
MAO/CEIL/M1 - missing window blower
DataStreams:maoceilpblhtM1.a0, maoceilM1.b1
Description:
The window blower was missing from the VCEIL at deployment and could not be found. A new 
unit was ordered and installed. Data may be affected during periods of high humidity, 
condensation, and precipitation.
Suggestions: 
Data are suspect when the window_contamination flag is set high.
Measurements:maoceilM1.b1:
  • first_cbh
  • second_cbh
  • third_cbh
  • vertical_visibility
more
maoceilpblhtM1.a0:
  • third_cbh
  • bl_height_1
  • bl_height_3
  • bl_height_2
  • first_cbh
  • second_cbh
more