DQRID : D140416.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/25/2014150003/26/20141215Suspect
more
Subject:
SGP/RL/C1 - Reduced SNR due to partial laser blockage
DataStreams:sgprlC1.a0
Description:
The laser output was partial blocked. This resulted in reduced SNR in all detection 
channels.
Suggestions: 
The data are useable but one may notice a slight increase in the noise.
Measurements:sgprlC1.a0:
  • t2_analog_high
  • water_counts_low
  • water_analog_low
  • depolarization_analog_high
  • nitrogen_counts_high
  • liquid_analog_high
  • t2_counts_high
  • depolarization_counts_high
  • elastic_counts_high
  • water_counts_high
  • t1_counts_high
  • nitrogen_analog_low
  • water_analog_high
  • elastic_analog_high
  • nitrogen_analog_high
  • liquid_counts_high
  • elastic_counts_low
  • elastic_analog_low
  • t1_analog_high
  • nitrogen_counts_low
more