Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/22/2014 | 1901 | 07/22/2014 | 2022 | Incorrect |
07/10/2014 | 0500 | 07/22/2014 | 1602 | Incorrect |
07/25/2014 | 1607 | 07/25/2014 | 1807 | Incorrect |
Subject: | MAO/CPC/S1 - Unknown cut-size affecting interpreted concentrations |
DataStreams: | maoaoscpcuS1.a1 |
Description: | During this time period, the condenser thermistor was not functioning, resulting in no temperature control in the condenser and an unknown instrument super-saturation. This results in an unknown cut-size for the instrument, which is critical for interpreting the size range over which the measured concentration applies. |
Suggestions: | Do not use data |
Measurements: | maoaoscpcuS1.a1: more
|