DQRID : D141124.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/05/2014130011/20/20142000Incorrect
more
Subject:
SGP/SIRS/E15 - Instrument problem
DataStreams:sgpsirsE15.b1, sgpsirs20sE15.a0
Description:
Short_direct_normal instrument not operating correctly during the affected time period.
Suggestions: 
NIP data are not recoverable.  Direct normal data can be reconstructed from global 

hemispheric and global diffuse data using the following relationship:
   global hemispheric = direct*cos(sza) + diffuse hemispheric
Measurements:sgpsirs20sE15.a0:
  • inst_direct_normal
more
sgpsirsE15.b1:
  • short_direct_normal
  • short_direct_normal_max
  • short_direct_normal_min
  • short_direct_normal_std
more