DQRID : D141204.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/10/2014151212/11/20141912Does not affect quality
11/29/2014120012/03/20141340Does not affect quality
more
Subject:
MAO/NEPHELOMETER/M1  - Dilution pump off from power outlet failure
DataStreams:maoaoshumidgrM1.00, maoaosnephwetM1.b1
Description:
The dilution pump was off due to problems in the trailer power. As a result the humidifier 
was turned off and was not scanning RH in order to prevent flooding in the wet neph.
Suggestions: 
Measurements:maoaoshumidgrM1.00:
  • Raw data stream - documentation not supported
more
maoaosnephwetM1.b1:
  • base_time
  • alt
  • qc_T_Neph_Inlet_Wet
  • Bs_B_Wet_Neph3W_2
  • Bs_G_Wet_Neph3W_2
  • qc_RH_Neph_Vol_Wet
  • Bbs_B_Wet_Neph3W_2
  • time_offset
  • lat
  • Bbs_G_Wet_Neph3W_2
  • qc_Bbs_G_Wet_Neph3W_2
  • P_Neph_Wet
  • T_Neph_Vol_Wet
  • qc_Bbs_B_Wet_Neph3W_2
  • qc_RH_Neph_Inlet_Wet
  • qc_T_Neph_Wet
  • qc_impactor_setting
  • Bbs_R_Wet_Neph3W_2
  • lon
  • qc_Bs_G_Wet_Neph3W_2
  • qc_Bs_R_Wet_Neph3W_2
  • qc_T_Neph_Vol_Wet
  • RH_Neph_Vol_Wet
  • impactor_setting
  • RH_Neph_Wet
  • Bs_R_Wet_Neph3W_2
  • time
  • qc_P_Neph_Wet
  • T_Neph_Inlet_Wet
  • qc_Bbs_R_Wet_Neph3W_2
  • qc_Bs_B_Wet_Neph3W_2
  • RH_Neph_Inlet_Wet
  • T_Neph_Wet
  • qc_RH_Neph_Wet
more

Close this window