Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/14/2014 | 1356 | 11/14/2014 | 1855 | Incorrect |
Subject: | MAO/AETH/NEPHELOMETER/PASS/PSAP/S1 - Impactor not reporting data |
DataStreams: | maoaosnephdryS1.a1, maoaospass3wS1.a1, maoaosaethS1.a1, maoaosaethS1.00, maoaospsap3wS1.b1, maoaospsap3wS1.a1, maoaosnephwetS1.b1, maoaosnephdryS1.b1, maoaospsap3wS1.01 |
Description: | After work to troubleshoot the AC sensor, DAQ box output to the impactor was hung and did not switch the impactor. For the time period there are some changes in the dryneph pressure. We cannot confirm that the impacter did not move out of the 10um position and should be considered unknown. |
Suggestions: | Do not use data for time period. |
Measurements: | maoaospass3wS1.a1: more
maoaospsap3wS1.a1: more maoaospsap3wS1.01: more maoaospsap3wS1.b1: more maoaosaethS1.a1: more maoaosaethS1.00: more maoaosnephdryS1.b1: more maoaosnephdryS1.a1: more maoaosnephwetS1.b1: more |