Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/07/2015 | 0000 | 07/09/2015 | 2130 | Incorrect |
Subject: | NSA/AERI/C1 - Detector temperature too high. |
DataStreams: | nsaaerich1C1.b1, nsaaerich2C1.a1, nsaaeriC1.00, nsaaerirawC1.00, nsaaeriengineerC1.b1, nsaaerisummaryC1.b1, nsaaeriengineerC1.a1, nsaaerich1C1.a1, nsaaerisummaryC1.a1, nsaaerich2C1.b1 |
Description: | During this time period the detector temperature was over 100 K and data collected is not valid. |
Suggestions: | Do not use radiances collected during this period. |
Measurements: | nsaaeriengineerC1.a1: more