Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/18/2015 | 0000 | 07/20/2015 | 2359 | Does not affect quality |
Subject: | SGP/AERI/C1 - Instrument run in slow sample mode. |
DataStreams: | sgpaerich2C1.a2, sgpaeri01ch2C1.a1, sgpaeriC1.00, sgpaeri01C1.00, sgpaerixC1.00, sgpaerisummaryC1.a1, sgpaeriengineerC1.b1, sgpaerisummaryC1.a2, sgpaeri01engineerC1.a1, sgpaerich1C1.b1, sgpaeri01summaryC1.a1, sgpaerisummaryC1.b1, sgpaerich1C1.a1, sgpaeri01ch1C1.a1, sgpaerich2C1.b1, sgpaerich2C1.a1, sgpaerich1C1.a2 |
Description: | Data is fine but the instrument was run in slow sample mode for these 3 days. We had changed the digital meter followed by a short 3rd body test that is run in slow sample mode. The instrument wasn't not switched back to rapid sample mode during this period. |
Suggestions: | Be aware that rather then records every 20 seconds there are records approximately every 8 minutes. |