Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/23/2015 | 1520 | 07/08/2015 | 1830 | Incorrect |
Subject: | MAO/SASHE/M1 - Instrument |
DataStreams: | maosashenirM1.b1, maosashevisM1.b1 |
Description: | Power supply for shutter failed preventing proper determination of background and proper determination of irradiance components. |
Suggestions: | Use selected MFRSR data if available. |
Measurements: | maosashevisM1.b1: more
maosashenirM1.b1: more |