DQRID : D160119.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/17/2015160011/18/20152207Incorrect
more
Subject:
SGP/AOS/C1 - Wet NEPH flooded
DataStreams:sgpaosnephwetC1.b1, sgpaosnephwetC1.a1
Description:
Following power outage wet NEPH flooded from back pressure in humidifier. Tech drained 
water and dried out instrument. Humidifier scanning resumed 01/16/2016 00:00 UTC after 
humidifier rebuilt and installed.
Suggestions: 
Do not use data
Measurements:sgpaosnephwetC1.b1:
  • impactor_setting
  • P_Neph_Wet
  • RH_Neph_Inlet_Wet
  • Bbs_G_Wet_Neph3W_2
  • Bbs_B_Wet_Neph3W_2
  • Bs_R_Wet_Neph3W_2
  • T_Neph_Vol_Wet
  • Bbs_R_Wet_Neph3W_2
  • Bs_B_Wet_Neph3W_2
  • RH_Neph_Vol_Wet
  • Bs_G_Wet_Neph3W_2
  • T_Neph_Inlet_Wet
more
sgpaosnephwetC1.a1:
  • RH_NephInlet_Wet
  • Bbs_G_Wet_10um_Neph3W_2
  • T_NephVol_Wet
  • Bbs_R_Wet_1um_Neph3W_2
  • Bbs_R_Wet_10um_Neph3W_2
  • Bbs_B_Wet_10um_Neph3W_2
  • Bs_G_Wet_10um_Neph3W_2
  • RH_NephVol_Wet
  • Bs_R_Wet_1um_Neph3W_2
  • Bbs_B_Wet_1um_Neph3W_2
  • Bs_R_Wet_10um_Neph3W_2
  • Bbs_G_Wet_1um_Neph3W_2
  • T_NephInlet_Wet
  • Bs_G_Wet_1um_Neph3W_2
  • P_Neph_Wet
  • Bs_B_Wet_10um_Neph3W_2
  • Bs_B_Wet_1um_Neph3W_2
more

Close this window