Subject: | ENA/OZONE/C1 - Reprocess: Under reporting ozone concentration |
DataStreams: | enaaoso3C1.00, enaaoso3C1.a1
|
Description: | Switching valve developed an internal leak that gradually worsened. Reported instrument
values are under-representations of the ambient ozone concentration during this period. The
under reporting worsens from 85% to 77% over the course of this DQR. No corrections have
been applied at this point. |
Suggestions: | A linear correction of the data from reported[ozone]/0.85 at the beginning of the
period to reported[ozone]/0.77 at the end of the period brings the internal calibrations
into line AND agrees with the step function of [O3] before and after the valve replacement. |
Measurements: | enaaoso3C1.a1: more
enaaoso3C1.00: - Raw data stream - documentation not supported
more
|