Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/14/2016 | 1930 | 02/09/2016 | 2030 | Suspect |
Subject: | SGP/RL/C1 - Suboptimal laser focus |
DataStreams: | sgprlC1.a0 |
Description: | The laser focus was less than optimal during this time period. Initially, it appeared that the focus was in too close. A first attempt at adjusting the focus was made at about 1830 UTC on 20160203. This resulted in the focus being adjusted too far out. The focus was then adjusted again to bring it back in a little at ~2030 on 20160209, and this did the trick. |
Suggestions: | Users are cautioned that the data recorded during the effected time are not optimal. |
Measurements: | sgprlC1.a0: more
|