DQRID : D160211.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/01/2016194502/11/20161625Incorrect
more
Subject:
OLI/DL/M1 - Instrument not scanning
DataStreams:olidlusrM1.b1, olidlppiM1.b1, olidlcal2M1.b1, olidlfptM1.b1
Description:
The elevation motor on the scanner was stuck at 0deg, thus all stares and scans (i.e. PPI) 
were done with elevation=0. Also the intensity and backscatter fields contain erroneous 
negative values during this time period.
Suggestions: 
Do not use the intensity or backscatter fields during this period.
Measurements:olidlfptM1.b1:
  • attenuated_backscatter
  • intensity
more