DQRID : D160218.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/12/2016144401/15/20161935Incorrect
more
Subject:
SGP/PSAP/C1 - PSAP flow incorrect
DataStreams:sgpaospsap3wC1.b1, sgpaospsap3w1mC1.b1, sgpaospsap3wC1.a1
Description:
PSAP flow was set too low causing noise and particle loss
Suggestions: 
USE CLAP data
Measurements:sgpaospsap3w1mC1.b1:
  • Ba_G_raw
  • Ba_R_raw
  • impactor_state
  • transmittance_green
  • Ba_G_Weiss
  • sample_volume
  • spot_size_area
  • Ba_R_Weiss
  • transmittance_blue
  • transmittance_red
  • Ba_B_raw
  • Ba_B_Weiss
  • sample_flow_rate
more
sgpaospsap3wC1.a1:
  • averaging_time
  • Ba_G_Dry_1um_PSAP3W_1
  • Ba_B_Dry_10um_PSAP3W_1
  • mass_flow_voltage
  • Ba_R_Dry_1um_PSAP3W_1
  • psap_flow_rate
  • Ba_B_Dry_1um_PSAP3W_1
  • Ba_B_PSAP3W
  • Ba_G_Dry_10um_PSAP3W_1
  • Ba_R_PSAP3W
  • TrG
  • TrB
  • impactor_setting
  • Ba_R_Dry_10um_PSAP3W_1
  • mass_flow
  • psap_flags
  • transmittance_B
  • filter_id
  • transmittance_R
  • Ba_G_PSAP3W
  • sample_length
  • transmittance_G
  • TrR
more
sgpaospsap3wC1.b1:
  • transmittance_green
  • absorbance_green
  • dilution_correction_factor
  • filter_id
  • absorbance_red
  • dilution_flow_setpoint
  • Ba_R_PSAP3W
  • Ba_B_PSAP3W
  • transmittance_blue
  • sample_flow_rate
  • dilution_flow
  • transmittance_red
  • transmittance_R
  • impactor_setting
  • transmittance_G
  • absorbance_blue
  • transmittance_B
  • dilution_flow_valve_position
  • spot_size_area
  • impactor_state
  • mass_flow
  • Ba_G_PSAP3W
  • seconds_after_transition
more

Close this window