Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/11/2016 | 0055 | 02/12/2016 | 1334 | Incorrect |
02/12/2016 | 2125 | 02/19/2016 | 1651 | Incorrect |
01/14/2016 | 2044 | 02/07/2016 | 1125 | Incorrect |
Subject: | ENA/MFRSR/C1 - Heater not working - connector issue |
DataStreams: | enamfrsrauxC1.a0, enamfrsrlangleyC1.c1, enamfrsrC1.b1, enamfrsrC1.00, enamfrsrC1.a0, enamfrsrlangplotC1.c1 |
Description: | There are three periods when the heater is not working due to a poor connection. Instrument is designed to operate at ~40 degC. Avoid data covered by this DQR. |
Suggestions: | Do not use data |
Measurements: | enamfrsrlangleyC1.c1: more