Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/17/2016 | 2300 | Open Ended | 2300 | Does not affect quality |
Subject: | SGP/CCN/C1 - Reprocess: CCN salt calibration change data processing |
DataStreams: | sgpaosccnavgC1.c1, sgpaosccn100C1.a1, sgpaosccnavgC1.c2, sgpaosccn1colC1.b1, sgpaosccnC1.a1, sgpaosccnC1.00 |
Description: | The CCN instrument calibration was changed to dT=16.148(SS)+0.8406. Data processing needs to change to incorporate this calibration into the model calculations |
Suggestions: | Wait for reprocessed data. |
Measurements: | sgpaosccnC1.a1: more
sgpaosccnC1.00: more sgpaosccn1colC1.b1: more sgpaosccn100C1.a1: more sgpaosccnavgC1.c2: more sgpaosccnavgC1.c1: more |