Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/17/2016 | 1600 | 03/18/2016 | 1800 | Does not affect quality |
Subject: | SGP/CCN/C1 - % super saturation values change after instrument calibration |
DataStreams: | sgpaosccnavgC1.c1, sgpaosccn100C1.a1, sgpaosccnavgC1.c2, sgpaosccn1colC1.b1, sgpaosccnC1.a1 |
Description: | % super saturation scan values during this time period were 0.12, 0.23, 0.4, 0.85, 0.51, 0.29 and 0.18. Before and after this time period the % super saturation scan values are 0.15, 0.25, 0.4, 0.8, 0.5, 0.3 and 0.2. |
Suggestions: | |
Measurements: | sgpaosccnC1.a1: more