DQRID : D160318.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/17/2016160003/18/20161800Does not affect quality
more
Subject:
SGP/CCN/C1 - % super saturation values change after instrument calibration
DataStreams:sgpaosccnavgC1.c1, sgpaosccn100C1.a1, sgpaosccnavgC1.c2, sgpaosccn1colC1.b1,
sgpaosccnC1.a1
Description:
% super saturation scan values during this time period were 0.12, 0.23, 0.4, 0.85, 0.51, 
0.29 and 0.18. Before and after this time period the % super saturation scan values are 
0.15, 0.25, 0.4, 0.8, 0.5, 0.3 and 0.2.
Suggestions: 
Measurements:sgpaosccnC1.a1:
  • CCN_ss_set
more
sgpaosccn1colC1.b1:
  • supersaturation_set_point
more
sgpaosccn100C1.a1:
  • CCN_ss_set
more
sgpaosccnavgC1.c2:
  • supersaturation_setpoint
more
sgpaosccnavgC1.c1:
  • supersaturation_setpoint
more

Close this window