Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/22/1995 | 0000 | 03/12/2014 | 2359 | Does not affect quality |
01/07/2011 | 0000 | 02/24/2016 | 0000 | Does not affect quality |
Subject: | SGP/AERI/C1 - Instrument |
DataStreams: | sgpaeri01summaryC1.a1, sgpaeri01engineerC1.a1, sgpaeri01ch2C1.a1, sgpaeri01ch1C1.a1, sgpaeri01C1.00 |
Description: | A problem was found with the new AERI in the optical trailer at the central facility. The climate controlled air from inside the optical trailer was penetrating into the AERI front-end under certain conditions, thereby contaminating the measurements in the atmospheric layer nearest the surface. Some cases have been found where the AERI-01 (adjacent to the new AERI) also showed this contamination. The symptoms are a ~15 min oscillation in observed radiance that correlates with the temperature inside the optical trailer. Note that data averaged over periods >~15 mins will be minimally affected. |
Suggestions: | The mentor team is developing an algorithm to identify the cases where this problem occurred. The signature is a correlation between the temperature inside the optical trailer (e.g. SCEtemp) and the temperature in the front-end (e.g. BBsupportStructureTemp) |
Measurements: | sgpaeri01ch1C1.a1: more
sgpaeri01summaryC1.a1: more sgpaeri01ch2C1.a1: more sgpaeri01engineerC1.a1: more sgpaeri01C1.00: more |