Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/05/2016 | 1720 | 05/10/2016 | 1604 | Incorrect |
Subject: | SGP/MFRSR/E35 - Instrument problem, ch 415 not functioning |
DataStreams: | sgpmfrsrE35.00, sgpmfrsrlangplotE35.c1, sgpmfrsrlangleyE35.c1, sgpmfrsrE35.b1, sgpmfrsrE35.a0, sgpmfrsrauxE35.a0 |
Description: | Channel 415 is not functioning during the period covered by this DQR. The problem was associated with the signal cable. The cable was replaced and 415 came back. The cable may be defective, or it's possible it was making a poor connection at the sensor. |
Suggestions: | |
Measurements: | sgpmfrsrlangplotE35.c1: more
sgpmfrsrE35.00: more sgpmfrsrlangleyE35.c1: more sgpmfrsrE35.a0: more sgpmfrsrE35.b1: more sgpmfrsrauxE35.a0: more |