DQRID : D160614.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/2016220006/14/20161357Incorrect
more
Subject:
SGP/CLAP/C1 - Dirty optics and leak in valve
DataStreams:sgpaosclap3wC1.b1
Description:
Leak in valve as well as dirty optics lowered data quality and made it unusable. 
Instrument was removed for repair and cleaning.
Suggestions: 
Do not use data. Use PSAP data.
Measurements:sgpaosclap3wC1.b1:
  • transmittance_red
  • transmittance_red_normalization_factor
  • Ba_B_CLAP3W_1
  • impactor_setting
  • impactor_state
  • transmittance_R
  • dilution_correction_factor
  • transmittance_blue_normalization_factor
  • filter_id
  • absorbance_red
  • transmittance_blue
  • transmittance_green_normalization_factor
  • seconds_after_transition
  • sample_flow_rate
  • transmittance_green
  • active_filter_number
  • transmittance_B
  • sample_length
  • Ba_R_CLAP3W_1
  • transmittance_G
  • absorbance_blue
  • Ba_G_CLAP3W_1
  • absorbance_green
  • spot_size_area
more

Close this window