Subject: | SGP/CCN/S01 - Concentration measurement differences |
DataStreams: | sgpaosccn200S01.00, sgpaosccn200S01.a1
|
Description: | Total particle number concentrations different approximately by 20-30% between the two
columns at the same supersaturation setpoints. Possible cause - differences in column
temperature gradients. Data is suspect. Updated calibration coefficients are needed to derive
correct supersaturation values from the recorded temperatures. |
Suggestions: | Supersaturation values need to be recalculated. |
Measurements: | sgpaosccn200S01.a1: - N_CCN_dN_2
- N_CCN_dN_1
- N_CCN_1
- CCN_supersaturation_set_point_1
- N_CCN_2
- CCN_supersaturation_set_point_2
more
sgpaosccn200S01.00: - Raw data stream - documentation not supported
more
|