DQRID : D160707.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/11/2016021507/18/20161500Incorrect
06/30/2016211007/01/20162359Incorrect
more
Subject:
SGP/CPC/C1 - Low CPC signal
DataStreams:sgpaoscpcC1.b1, sgpaoscpcC1.a1
Description:
The CPC signal was low as the result of a weak laser. As TSI no longer services this 
instrument no replacement is available.
Suggestions: 
Do not use data
Measurements:sgpaoscpcC1.b1:
  • f2_flags
  • dilution_correction_factor
  • concentration
  • cpc_flow
more
sgpaoscpcC1.a1:
  • concentration
  • laser_current
  • condensation_temp
  • ambient_temp
  • nozzle_pressure
  • optics_temp
  • cabinet_temp
  • orifice_pressure
  • f2_flags
  • saturation_temp
  • cpc_flow
more

Close this window