Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/2016 | 1300 | 05/23/2016 | 1505 | Does not affect quality |
Subject: | ASI/AOS/M1 - Impactor state not read |
DataStreams: | asiaosimpactorM1.b1, asiaosimpactorM1.a1, asiaosimpactorM1.00 |
Description: | Impactor state was not read back from the limit switches. Impactor still switching and the impactor state data is valid. This applies only to raw data as the read-back state is only in raw data. |
Suggestions: | When using RAW data, use the impactor state set data and note the values read back from the impactor. |
Measurements: | asiaosimpactorM1.b1: more
asiaosimpactorM1.a1: more asiaosimpactorM1.00: more |