Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/14/2016 | 1200 | 06/14/2016 | 2359 | Suspect |
Subject: | NSA/MET/C1 - Chilled mirror hygrometer noise |
DataStreams: | nsametC1.b1, nsametC1.00 |
Description: | During this time, the chilled mirror hygrometer experiences periods of noisy humidity data, likely due to mirror contamination. |
Suggestions: | Use dew_point_mean, rh_mean, and vapor_pressure_mean from this data stream as an alternative. |
Measurements: | nsametC1.b1: more
nsametC1.00: more |