DQRID : D160715.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/11/2016194004/11/20162049Incorrect
04/11/2016190104/11/20161918Incorrect
04/11/2016172704/11/20161743Incorrect
06/21/2016183906/21/20162146Incorrect
06/24/2016164506/24/20161651Incorrect
04/12/2016193204/12/20162026Incorrect
04/11/2016153604/11/20161608Incorrect
04/12/2016153904/12/20161713Incorrect
more
Subject:
SGP/MFRSR/E34 - Instrument Noise, ch615
DataStreams:sgpmfrsrlangleyE34.c1, sgpmfrsrlangplotE34.c1, sgpmfrsrE34.00, sgpmfrsrauxE34.a0,
sgpmfrsrE34.b1, sgpmfrsrE34.a0
Description:
There are a few periods where channel 615 has noise or drops out completely.  We believe 
the problem was related to a bad connection.  The sensor cable was replaced on 20160629.  
Since the maintenance there hasn't been any issue with 615, and we're now hopeful the 
issue has been resolved.
Suggestions: 
Measurements:sgpmfrsrE34.a0:
  • alltime_hemisp_narrowband_filter3
  • hemisp_narrowband_filter3
  • hemisp_narrowband_filter3_raw
  • direct_horizontal_narrowband_filter3
  • diffuse_hemisp_narrowband_filter3
  • diffuse_hemisp_narrowband_filter3_raw
more
sgpmfrsrE34.b1:
  • direct_normal_narrowband_filter3
  • hemisp_narrowband_filter3
  • diffuse_hemisp_narrowband_filter3
  • alltime_hemisp_narrowband_filter3
  • direct_diffuse_ratio_filter3
  • direct_horizontal_narrowband_filter3
more
sgpmfrsrlangplotE34.c1:
  • barnard_lnI_filter3
  • michalsky_lnI_filter3
more
sgpmfrsrlangleyE34.c1:
  • barnard_optical_depth_filter3
  • michalsky_optical_depth_filter3
more
sgpmfrsrE34.00:
  • Raw data stream - documentation not supported
more
sgpmfrsrauxE34.a0:
  • blk_filter3
  • fsb_filter3
  • ssb_filter3
  • th_filter3
more

Close this window