Description: | The v-channel velocity (uncorrected_mean_doppler_velocity_v) is incorrectly filled with
NaNs. H-channel is correct.
The receiver gain (r_calib_receiver_gain_hc and r_calib_receiver_gain_vc) is incorrect.
This does not affect the data quality.
The incorrect v-channel calibration constant (r_calib_radar_constant_v) is reported in the
file. The correct value should be: -23.1151218
Similarly, the reported signal-to-noise ratio (signal_to_noise_ratio_cross_polar_v) is
calculated using the incorrect v-channel calibration constant, and is biased by a constant
value of -0.1999974
The reflectivity is calculated with the correct calibration constant and is unaffected. |