Subject: | ENA/MFRSR/C1 - Heater Problem, connector issue |
DataStreams: | enamfrsrauxC1.a0, enamfrsrlangleyC1.c1, enamfrsrC1.b1, enamfrsrC1.00, enamfrsrC1.a0, enamfrsrlangplotC1.c1 |
Description: | It seems like there may be a problem with the signal connector. We've had a few instances of the head temp failing, but replacing the heater board did not resolve the problem. For the most recent "fix" the system was powered down and the signal cable reseated. |
Suggestions: | |
Measurements: | enamfrsrlangleyC1.c1: more
enamfrsrlangplotC1.c1: more enamfrsrC1.b1: more enamfrsrC1.a0: more enamfrsrC1.00: more enamfrsrauxC1.a0: more |