DQRID : D160906.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/14/2016110008/25/20161459Suspect
more
Subject:
SGP/IRSI/C1 - Cooler failure
DataStreams:sgpirsiirC1.b1
Description:
The cooling device failed allowing the interior of the enclosure to consistently exceed 40 
C. An external air conditioner was powered-on, a solar-reflective panel was added, and 
the internal cooler was replaced.
Suggestions: 
Use data with caution
Measurements:sgpirsiirC1.b1:
  • sky_cover_high_emission_narrow
  • sky_cover_high_emission_wide
  • precipitable_water_vapor
  • sky_cover_low_emission_wide
  • sky_cover_low_emission_narrow
more

Close this window