Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/12/2016 | 2055 | 08/31/2016 | 1249 | Incorrect |
Subject: | SGP/ACSM/C1 - Filter switching valve not operating |
DataStreams: | sgpaosacsmC1.00, sgpaosacsmC1.01, sgpaosacsmC1.a1, sgpaosacsmcalC1.b1, sgpaosacsmC1.b1 |
Description: | Filter switching valve was not operating because of a faulty circuit board. Problem was fixed and data have been collected from the end of the specified time period. |
Suggestions: | Data are incorrect and should not be used. |
Measurements: | sgpaosacsmC1.01: more
sgpaosacsmcalC1.b1: more sgpaosacsmC1.a1: more sgpaosacsmC1.00: more sgpaosacsmC1.b1: more |