Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/17/2016 | 1035 | 08/18/2016 | 0538 | Incorrect |
08/15/2016 | 0202 | 08/15/2016 | 0939 | Incorrect |
08/20/2016 | 1348 | 08/21/2016 | 1726 | Incorrect |
08/24/2016 | 0840 | 08/24/2016 | 1105 | Incorrect |
08/01/2016 | 0000 | 08/04/2016 | 1832 | Incorrect |
Subject: | ENA/OZONE/C1 - Water condensing in the line |
DataStreams: | enaaoso3C1.01, enaaoso3C1.00, enaaoso3C1.a1 |
Description: | Instrument suffered major oscillations in the signal. Working with Operator it was discovered that the Split System Air Conditioner was blowing directly on the Ozone detector. Water was condensing in the line. The air was redirected and the temperature setpoint raised. This reduced but didn't entirely eliminate the oscillations. |
Suggestions: | Do not use data |
Measurements: | enaaoso3C1.01: more
enaaoso3C1.a1: more enaaoso3C1.00: more |