DQRID : D160909.17
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/17/2016103508/18/20160538Incorrect
08/15/2016020208/15/20160939Incorrect
08/20/2016134808/21/20161726Incorrect
08/24/2016084008/24/20161105Incorrect
08/01/2016000008/04/20161832Incorrect
more
Subject:
ENA/OZONE/C1 - Water condensing in the line
DataStreams:enaaoso3C1.01, enaaoso3C1.00, enaaoso3C1.a1
Description:
Instrument suffered major oscillations in the signal.  Working with Operator it was 
discovered that the Split System Air Conditioner was blowing directly on the Ozone detector.  
Water was condensing in the line.  The air was redirected and the temperature setpoint 
raised.  This reduced but didn't entirely eliminate the oscillations.
Suggestions: 
Do not use data
Measurements:enaaoso3C1.01:
  • ozone
  • ozone_bench_temp
  • ozone_pressure
more
enaaoso3C1.a1:
  • ozone_bench_temp
  • ozone
  • ozone_pressure
more
enaaoso3C1.00:
  • Raw data stream - documentation not supported
more