DQRID : D160919.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/03/2016165507/21/20162359Suspect
more
Subject:
ASI/SASZE/M1 - Spectrometer chiller temperature variations
DataStreams:asisaszefilterbandsM1.a1, asisaszevisM1.a1
Description:
Spectrometer chiller temperature variations caused by an incorrectly placed sensor are 
apparent as correlated modulations in the signal from the UV-VIS CCD spectrometer. The 
magnitude of signal oscillation is small and is only apparent under clear sky for wavelengths 
< 450 nm. The variations were reduced to negligible levels by repositioning the sensor on 
Aug. 21, 2016.
Suggestions: 
Either avoid use of zenith irradiance < 450 nm during clear sky or apply a temporal 

smoothing function.