Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2016 | 1132 | 10/04/2016 | 1623 | Suspect |
Subject: | ASI/ACSM/M1 - Critical orifice clogged |
DataStreams: | asiaosacsmM1.a1, asiaosacsmM1.b1, asiaosacsmM1.00 |
Description: | The critical orifice was clogged during the time period specified. The orifice was cleaned and the instrument placed back in service. |
Suggestions: | The data can be corrected for pressure by scaling the signal to the strength of the air beam, however, because of the low pressure, the signal will be low and potentially noisy. The data during the time period specified should be used with caution. |
Measurements: | asiaosacsmM1.b1: more
asiaosacsmM1.a1: more asiaosacsmM1.00: more |