Subject:
ASI/ACSM/M1 - Critical orifice clogged
DataStreams:asiaosacsmM1.a1, asiaosacsmM1.b1, asiaosacsmM1.00
Description:
The critical orifice was clogged during the time period specified.  The orifice was 
cleaned and the instrument placed back in service.
Suggestions: 
The data can be corrected for pressure by scaling the signal to the strength of the 

air beam, however, because of the low pressure, the signal will be low and potentially 
noisy.  The data during the time period specified should be used with caution.