Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/06/2016 | 2036 | 10/11/2016 | 2137 | Does not affect quality |
Subject: | ASI/MPL/M1 - Instrument averaging time increased to 30 seconds |
DataStreams: | asimplpolfsM1.b1, asimplpolfsM1.00 |
Description: | To understand what was causing frequent spikes in the MPL energy levels, the averaging time was increased to 30s during this period. The averaging time was reverted back to the original 10s averaging time at the end of the DQR period. |
Suggestions: | |
Measurements: | asimplpolfsM1.b1: more