DQRID : D161130.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/20/2016020011/21/20161700Incorrect
11/16/2016133011/17/20161600Incorrect
11/22/2016170011/24/20161600Incorrect
more
Subject:
AWR/HTDMA/M1 - Instrument connectivity and flow issues
DataStreams:awraoshtdmaM1.a1, awraoshtdmaM1.00
Description:
HSEMS intermittently losing connection with SEMS and not setting the classifier diameter. 
MCPC not drawing in butanol due to internal pressure imbalance. Data incorrect.
Suggestions: 
Do not use the data from these time periods
Measurements:awraoshtdmaM1.a1:
  • scan_min_diameter_setting
  • dry_diameter_setting
  • num_bins
  • scan_type
  • humid_rh
  • total_concentration
  • scan_time_setting
  • sample_flow
  • dry_rh_setting
  • bin_width
  • aerosol_concentration
  • bin_center
  • atomizer
  • bin
  • scan_max_diameter_setting
more
awraoshtdmaM1.00:
  • Raw data stream - documentation not supported
more

Close this window