Subject: | AWR/HTDMA/M1 - Instrument connectivity and flow issues |
DataStreams: | awraoshtdmaM1.a1, awraoshtdmaM1.00
|
Description: | HSEMS intermittently losing connection with SEMS and not setting the classifier diameter.
MCPC not drawing in butanol due to internal pressure imbalance. Data incorrect. |
Suggestions: | Do not use the data from these time periods |
Measurements: | awraoshtdmaM1.a1: - scan_min_diameter_setting
- dry_diameter_setting
- num_bins
- scan_type
- humid_rh
- total_concentration
- scan_time_setting
- sample_flow
- dry_rh_setting
- bin_width
- aerosol_concentration
- bin_center
- atomizer
- bin
- scan_max_diameter_setting
more
awraoshtdmaM1.00: - Raw data stream - documentation not supported
more
|