DQRID : D161209.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/27/2016000012/02/20160000Does not affect quality
more
Subject:
OLI/TRACEGAS/M1 - Weekly displacement calibration of CO did not occur
DataStreams:oliaoscoM1.a1, oliaoscoM1.00, oliaoscoM1.b1
Description:
Some (minor) uncertainty in the calibration over this period.  However, Calibrations 
before and after the event appear very stable.  There are also daily standard additions 
between the weekly displacement calibrations.  These daily checks are quite consistent.  So 
this DQR is merely to notify of the condition.  Data can be used with confidence in the 
judgment of the Mentor.
Suggestions: 
Measurements:oliaoscoM1.b1:
  • seconds_after_calibration
  • ambient_temperature
  • mass_flow_through_MFC_2
  • mass_flow_through_MFC_1
  • co_dry
  • n2o
  • co
  • set_point_for_MFC_2
  • h2o
  • n2o_dry
  • set_point_for_MFC_1
  • valve_position_MFC_1
  • valve_position_MFC_2
  • gas_temperature
  • gas_pressure
more
oliaoscoM1.a1:
  • h2o
  • set_point_for_MFC_2
  • set_point_for_MFC_1
  • gas_temperature
  • mass_flow_through_MFC_2
  • ambient_temperature
  • mass_flow_through_MFC_1
  • gas_pressure
  • n2o_dry
  • valve_position_MFC_1
  • co_dry
  • co
  • valve_position_MFC_2
  • n2o
more
oliaoscoM1.00:
  • Raw data stream - documentation not supported
more

Close this window