DQRID : D161215.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/15/2016120012/15/20161832Does not affect quality
more
Subject:
ENA/OZONE/C1 - missing data and extra calibrations
DataStreams:enaaoso3C1.01, enaaoso3C1.00, enaaoso3C1.a1
Description:
The 49i ozone instrument in ENA has difficulty with maintaining the data connection and 
the data stream while receiving the commands to do zeros and calibrations and has had gaps 
in the data that interfere with data processing.  During the time range a new version of 
the software that should be better at recovering from the errors was being tested.  There 
is a 2 hour gap in the data and a section of time with zeros and cals every 20 minutes.
Suggestions: 
Measurements:enaaoso3C1.01:
  • ozone
  • ozone_bench_temp
  • ozone_pressure
more
enaaoso3C1.a1:
  • ozone_bench_temp
  • ozone
  • ozone_pressure
more
enaaoso3C1.00:
  • Raw data stream - documentation not supported
more

Close this window