Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/01/2016 | 0000 | 12/12/2016 | 1740 | Suspect |
Subject: | SGP/MFRSR/E11 - Heater Board Failure |
DataStreams: | sgpmfrsrlangplotE11.c1, sgpmfrsrauxE11.a0, sgpmfrsrlangleyE11.c1, sgpmfrsrE11.00, sgpmfrsrE11.a0, sgpmfrsrE11.b1 |
Description: | The heater began showing signs of failure of 20161201. This presented as excessive noise in the data, predominately on the 1st, 5th and 7th. The heater board was replaced on 20161212, but unfortunately it went in with a blown fuse. This is covered in D170123.3. |
Suggestions: | |
Measurements: | sgpmfrsrlangplotE11.c1: more
sgpmfrsrE11.00: more sgpmfrsrlangleyE11.c1: more sgpmfrsrE11.a0: more sgpmfrsrE11.b1: more sgpmfrsrauxE11.a0: more |