Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/19/1997 | 0000 | 08/30/1997 | 2359 | Suspect |
Subject: | TWP/SKYRAD/C1 - NIP: Moisture inside NIP |
DataStreams: | twpskyrad60sC1.b1 |
Description: | The NIP took on a significant amount of moisture during a rainy period that ended on 970719. From this date until the RESET visit in August, the NIP data is corrupted. The errors are large (often greater than 100 W/m^2 in the direct beam) in the morning but tend to converge to the expected values in the afternoon. In general, NIP data from this period should be used with extreme caution. |
Suggestions: | use with extreme caution, much of the data are corrupt |
Measurements: | twpskyrad60sC1.b1: more
|