Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/18/2002 | 0500 | 02/08/2002 | 0400 | Suspect |
Subject: | TWP/CEIL/C1 - Reprocess: Incorrect units on installed ceilometer |
DataStreams: | twpceilC1.b1 |
Description: | The units setting on the ceilometer installed on 18 Jan. 2002 were incorrectly set to feet rather than meters. |
Suggestions: | The units are being reset. All heights should be multiplied by 0.3048 during this period. |
Measurements: | twpceilC1.b1: more
|