Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/18/2002 | 0132 | 05/16/2002 | 1830 | Suspect |
Subject: | NSA/AERI/C1 - AERI extended range detector at Barrow |
DataStreams: | nsaaerisummaryC1.a1, nsaaeriengineerC1.a1, nsaaerich1C1.a1, nsaaerich2C1.a1 |
Description: | The extended range detector had failed and a borrowed replacement detector was installed. The replacement detector had enough air present in the dewar to cause the cryogenic cooler to work at full current. The detector was still warmer than optimal and there is more noise present in the data than is considered normal. The data must be used with discretion. The system was shut off on 020516 due to noise and rising detector temperatures. |
Suggestions: | |
Measurements: | nsaaeriengineerC1.a1: more
nsaaerich1C1.a1: more nsaaerisummaryC1.a1: more nsaaerich2C1.a1: more |