DQRID : D021201.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/21/2002113011/21/20021311Suspect
more
Subject:
SGP/SONDE/C1 - Questionable RH CF sounding 0211211:1130
DataStreams:sgpsondewnpnC1.b1, sgpsondewnpnC1.a1
Description:
The RH samples in this sounding oscillate between reasonable values (60-80%) and zero.  
This may be a problem with the RH sensor or the software that alternates between the heated 
and non-heated RH elements.  In either case, the reported values are questionable.  More 
investigation of the raw binary data would be required to diagnose further.
Suggestions: 
No fix
Measurements:sgpsondewnpnC1.b1:
  • rh
  • dp
more
sgpsondewnpnC1.a1:
  • rh
  • dp
more