DQRID : D030317.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/27/2003175003/03/20031900Incorrect
more
Subject:
TWP/SMET/C1 - Reprocess: Lower Wind Speed incorrect
DataStreams:twpsmet60sC1.a1, twpsmet60sC1.b1
Description:
The lower wind speed sensor (sensor #2) was reporting values 10 times above the actual 
value.  During an update to the logger configuration file the calibration multiplier was 
incorrectly entered.  The multiplier inserted was .97630 when it should have been .09763.  
This caused the readings from this sensor to be incorrect.
Suggestions: 
Use upper wind speed (sensor #1) data or wait for reprocessed data.