DQRID : D031216.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/11/2003190610/15/20030015Missing
more
Subject:
NSA/GNDRAD/C1 - IRT returned to vendor
DataStreams:nsagndrad60sC1.b1
Description:
The IRT was returned to the vendor for diagnosis. It was found to be operating as expected 
and the signal noise problem was traced to the cable. The cable was repaired and the 
instrument reinstalled.
Suggestions: 
Measurements:nsagndrad60sC1.b1:
  • sfc_ir_temp
  • sfc_ir_temp_std
  • sfc_ir_temp_min
  • sfc_ir_temp_max
more