DQRID : D051020.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/27/2005220009/30/20052000Suspect
more
Subject:
NSA/MWR/C1 -  Incorrect time stamp
DataStreams:nsamwrlosC1.b1, nsamwrC1.00
Description:
On Sept 27 the motherboard of the MWR computer was replaced. Following the replacement, 
the date on the computer was not set to the correct time of day, therefore the date and 
time in the files were wrong. The time on the computer was reset on Sept. 30. Time stamps 
between Sept. 27 and Sept. 30 were later corrected based on gathered information, however 
they may have as much as 5 minutes of uncertainty.
Suggestions: 
You should not use these data at the highest temporal resolution because there may be 

an offset. If you are time averaging, you should probably average at least over a 
5-minute interval
Measurements:nsamwrC1.00:
  • Raw data stream - documentation not supported
more
nsamwrlosC1.b1:
  • liq
  • actaz
  • time_offset
  • bb31
  • tkxc
  • tknd
  • lat
  • tnd_nom23
  • tbsky31
  • actel
  • tnd23
  • vap
  • sky31
  • alt
  • bbn23
  • tkair
  • sky23
  • sky_ir_temp
  • tc31
  • tbsky23
  • tnd_nom31
  • wet_window
  • tc23
  • lon
  • ir_temp
  • tkbb
  • base_time
  • bb23
  • bbn31
  • tnd31
more