DQRID : D060718.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/12/2006110007/21/20061635Incorrect
more
Subject:
SGP/SIRS/E9 - Tracker Failure
DataStreams:sgpsirs20sE9.a0, sgpsirsE9.00, sgpsirsE9.b1
Description:
On July 12th the SIRS tracker failed to wake up that morning.  On the 21st of July CM 
repaired the tracker.
Suggestions: 
Data from tracker mounted instruments:NIP(short direct normal) and 8-48 (down short 

diffuse hemispheric) is incorrect and cannot be recovered.  The PIR (down long 
hemispheric shaded) data is unaffected.
Measurements:sgpsirsE9.b1:
  • down_short_diffuse_hemisp_std
  • down_short_diffuse_hemisp_min
  • short_direct_normal_min
  • down_short_diffuse_hemisp_max
  • short_direct_normal
  • down_short_diffuse_hemisp
  • short_direct_normal_std
  • short_direct_normal_max
more
sgpsirs20sE9.a0:
  • inst_diffuse
  • inst_direct_normal
more
sgpsirsE9.00:
  • Raw data stream - documentation not supported
more

Close this window