DQRID : D070306.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/03/2006150903/06/20071610Does not affect quality
more
Subject:
SGP/SIRS/C1 -  Reprocessed and Corrected: Incorrect calibration coefficients
DataStreams:sgpsirsC1.00, sgpsirsC1.b1, sgpsirs20sC1.a0
Description:
DATA REPROCESSED and CORRECTED: 

Incorrect calibration coefficients were used in the C1/SIRS data logger program. The 
PSP-US (upwelling shortwave global hemispheric), 30943F3 should be 126.93 instead Of 129.93 
[W/m2*mV]. The data were recalibrated to correct the error and reprocessed.

The data were archived in September 2016. Please retrieve the reprocessed data for use in 
your analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::sirs/fsite::sgp/ffac::sgp.E1 


ORIGINAL DQR CATEGORY: Red( Incorrect)

REPROCESSED and CORRECTED DATA ARCHIVED:2016-09

REPROCESS IDENTIFICATION NUMBER: RID 386

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------


Incorrection calibration coefficients were used in the C1/SIRS datalogger program. The 
PSP-US (upwelling shortwave global hemispheric), 30943F3 should be 126.93 not 129.93 
[W/m2/mV].
Suggestions: 
Measurements:sgpsirsC1.b1:
  • up_short_hemisp
  • up_short_hemisp_max
  • up_short_hemisp_min
  • up_short_hemisp_std
more
sgpsirsC1.00:
  • Raw data stream - documentation not supported
more
sgpsirs20sC1.a0:
  • inst_up_short_hemisp
more