DQRID : D070320.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/06/2006172204/12/20071810Does not affect quality
more
Subject:
SGP/SIRS/E22 - Reprocessed and Corrected: Incorrect calibration coefficients
DataStreams:sgpsirs20sE22.a0, sgpsirsE22.b1, sgpsirsE22.00
Description:
DATA REPROCESSED and CORRECTED: 

NIP (short_direct_normal) 29541E6 had an incorrect calibration coefficient in the logger 
program. The correct calibration should have been 126.67 [W/m2/mV] instead of 128.97. The 
correct calibration was calculated as NIPIrr(cor) = NIPIrr(orig) * (126.67/128.97) = 
NIP(orig) * 0.982166

The data were archived in October 2016. Please retrieve the reprocessed data for use in 
your analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::sirs/fsite::sgp/ffac::sgp.E22


ORIGINAL DQR CATEGORY:  Red (Incorrect)

REPROCESSED and CORRECTED DATA ARCHIVED:2016-10

REPROCESS IDENTIFICATION NUMBER: RID 387

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------

NIP (short_direct_normal) 29541E6 had an incorrect calibration coefficient in the logger 
program. 128.97 should have been 126.67 [W/m2/mV].
Suggestions: 
Measurements:sgpsirsE22.00:
  • Raw data stream - documentation not supported
more
sgpsirsE22.b1:
  • short_direct_normal
  • short_direct_normal_max
  • short_direct_normal_std
  • short_direct_normal_min
more
sgpsirs20sE22.a0:
  • inst_direct_normal
more

Close this window