DQRID : D070618.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/03/2007000006/26/20072025Suspect
more
Subject:
SGP/SMOS/E1 - Failed temperature probe
DataStreams:sgp30smosE1.b1, sgp1440smosE1.b1, sgp1smosE1.b1
Description:
The temperature probe was out of tolerance as it failed the 6-month calibration checks and 
subsequent bi-weekly checks.  The temperature values consistently were outside the 
manufacturers suggested uncertainty of 0.2 C.
Suggestions: 
Use temperature data from other nearby SMOS sites.
Measurements:sgp30smosE1.b1:
  • temp
  • vap_pres
  • sd_vap_pres
more
sgp1440smosE1.b1:
  • max_vap_pres
  • time_max_temp
  • min_vap_pres
  • min_temp
  • time_min_temp
  • time_min_vap_pres
  • time_max_vap_pres
  • max_temp
more
sgp1smosE1.b1:
  • temp
  • vap_pres
more

Close this window